Three methods for calibrating temperature sensors: pros and cons
There are three common methods for calibrating temperature sensors:
- Calibrating just the electronics with a simulator
- Calibrating both the electronics and the sensor in a dry-well
- Calibrating both the electronics and the sensor in a dry-well with a reference thermometer
Here is a quick look at these methods and the pros and cons of each.
Calibrating just the electronics with a simulator
To calibrate the electronics of a temperature measurement/control system, a temperature simulator takes an input temperature and outputs (simulates) the corresponding voltage (thermocouple) or resistance (RTD or thermistor) based upon accepted national tables.
Pros of this method:
- It’s fast. The electrical settings are instant and require no stabilization times.
- The calibration equipment may be more portable than the equipment used in other methods.
- The meter can still be used with interchangeable probes.
Cons of this method:
- You’ll need a separate calibration procedure for the probe or…
- The probe remains uncalibrated.
- Therefore, the thermometer system remains uncalibrated and measurements made with it are untraceable.
Calibrating both the electronics and the sensor in a dry-well
In this method, the dry-well is set to the desired test temperature and the thermometer, meter and probe combined, is placed in the well. The thermometer reading is compared to the reading on the dry-well’s built-in thermometer.
Pros of this method:
- The probe and meter are calibrated as a system.
- The actual temperature is used.
- System setup is simple as only a single instrument (the dry-well) is used.
Cons of this method:
- Accuracy is limited by the calibrated accuracy of the dry-well.
- If the probe does not reach the bottom of the well, additional error can occur.
- Unit-under-test probes are not interchangeable after the calibration.
- The dry-well requires time to change temperatures.
Calibrating both the electronics and the sensor in a dry-well with a reference thermometer
In this method, a dry-well is used as a stable heat source and the thermometer is compared against a more accurate thermometer placed in an adjacent hole in the well.
Pros of this method:
- This method has the best accuracy results.
- Calibration is traceable to the separate reference.
- You have more versatility in matching unit-under-test sizes.
Cons of this method:
- It’s more expensive than the other two methods.
- Because two instruments are used, setup is slightly more complex.
- The probes under test are not interchangeable after calibration.
- The dry-well requires time to change temperatures. Want to know more?
Conclusion
There are three primary methods for calibrating temperature sensors, and each method has distinct pros and cons. In summary, these methods are:
- Calibrating just the electronics with a simulator – This method is quick, but you’ll need a separate calibration procedure for the probe.
- Calibrating both the electronics and the sensor in the dry-well – This method allows you to calibrate the probe and meter as a system, but the accuracy will be limited by the calibrated accuracy of the dry-well.
- Calibrating both the electronics and the sensor in a dry-well with a reference thermometer – This method provides the best accuracy, but it is more expensive and complex than the other two methods.
Would you like to learn more about calibrating temperature sensors? Fluke Calibration offers free, online training in the form of on-demand webinars. This webinar expands upon the material covered in this blog post:
Additional resources
Thermocouple application note series >>
Temperature measurement and calibration resource center >>
Related products
Probes / Sensors (PRTs, Thermistors, and Thermocouples) >>
Digital Thermometer Readouts (DRTs) >>
Industrial Temperature Calibrators (Dry-Wells, Metrology Wells, and More) >>
- Inicie sesión o regístrese para enviar comentarios
- Printer-friendly version »
- Inicio
- Productos
- Novedades de Fluke Calibration
- Calibración eléctrica
- Calibración de radiofrecuencia
- Equipo de adquisición de datos y pruebas
- Calibración de temperatura
- Calibración de humedad
- Calibración de presión
- Manómetros de pistón
- Controladores/Calibradores
- Monitores de presión
- Balanzas de pesos muertos
- Monitores y calibradores manuales
- Handheld Pressure Calibrators
- Calibración de datos de aire
- Monitores ambientales
- Accesorios de calibración de presión
- Sistemas personalizados de calibración de presión
- Software de calibración de presión
- Calibración de flujo
- Calibradores de procesos
- Software de calibración
- Servicio y soporte técnico
- Todos los instrumentos de calibración
- Comprar
- Novedades
- Capacitación y eventos
- Literatura y educación
- Soporte Técnico
- Service Request (RMA)
- Programas de servicio
- Servicio técnico
- Base de Conocimientos
- Acreditaciones
- Centros de servicio autorizados
- Calibration Certificates
- Community Forum
- My MET/SUPPORT
- Manuales de productos (guías de usuario)
- Hojas de datos sobre la seguridad (SDS)
- Safety, Service, and Product Notices
- Descargas de software
- Sobre Nosotros