Q. Can I use MET/TEMP II to perform calibrations using fixed-point cells?
A. Yes. MET/TEMP II now supports using fixed-point cells. Because fixed-point cells typically allow a single probe to be inserted at a time, MET/TEMP II prompts you at the appropriate times to insert one UUT into the cell, counts down a user-defined soak time, then takes readings from that UUT. This process repeats for all UUTs.
Q. Can I mix fixed-points set-points and comparison set-points on the same calibration?
A. Fixed-point set-points can be mingled with comparison set-points, or MET/TEMP II can be configured to perform fixed-point only calibrations.
Q. What fixed-point cells can be used with MET/TEMP II?
A. MET/TEMP II ships with a list of all fixed-point cells that Hart manufactures. You are simply required to enter the serial number of your cell before it can be used. You can add support for as many fixed-point cells as you want. Simply fill in the model, serial number, manufacturer, description, method of realization, reference temperature and dates (optional) and add it to your list. Since MET/TEMP II cannot actually differentiate between a fixed-point cell and a cup of ice, this feature can prove very useful for adding support for many types of fixed temperature heat sources such as the Hart Model 9101 Ice Point or a bath or furnace that constantly operates at one temperature.
- Inicio
- Productos
- Novedades de Fluke Calibration
- Calibración eléctrica
- Calibración de radiofrecuencia
- Equipo de adquisición de datos y pruebas
- Calibración de temperatura
- Calibración de humedad
- Calibración de presión
- Manómetros de pistón
- Controladores/Calibradores
- Monitores de presión
- Balanzas de pesos muertos
- Monitores y calibradores manuales
- Handheld Pressure Calibrators
- Calibración de datos de aire
- Monitores ambientales
- Accesorios de calibración de presión
- Sistemas personalizados de calibración de presión
- Software de calibración de presión
- Calibración de flujo
- Calibradores de procesos
- Software de calibración
- Servicio y soporte técnico
- Todos los instrumentos de calibración
- Comprar
- Novedades
- Capacitación y eventos
- Literatura y educación
- Soporte Técnico
- Service Request (RMA)
- Programas de servicio
- Servicio técnico
- Base de Conocimientos
- Acreditaciones
- Centros de servicio autorizados
- Calibration Certificates
- Community Forum
- My MET/SUPPORT
- Manuales de productos (guías de usuario)
- Hojas de datos sobre la seguridad (SDS)
- Safety, Service, and Product Notices
- Descargas de software
- Sobre Nosotros